Spectrometer SM 9000

SM9000 SpectrometerThe new Spectrometer SM 9000 was developed for high-resolution, fiber optic spectroscopy applications in the UV/VIS to NIR range. It can be used as a stand-alone instrument or as a substantial enhancement of other PSI instruments to analyze spectral composition of irradiance. Ultra-high sensitivity of the instrument allows measuring fluorescence emission spectra of individual cells when combined with µ-FluorCam. The spectra can be measured in the range from UV (200 nm) to NIR (980 nm) with FWHM bandwidth of 3.5 nm. Fast recording can capture as many as 100 spectra per second with data resolution of 16 bits.

  • Ultra-high sensitivity.
  • Suitable for microspectrophotometry.
  • FWHM bandwidth of 3.5 nm.
  • Small-footprint, modular, robust, and thermally stable.
  • Very low heat generation.
  • 10 µs - 10 ms duration of flashes.
  • Variable exposure time, from 1 millisecond to minutes.
  • Kinetic spectra (max. 100 per second).
Technical specification:
  • Optical Entrance
    diameter 0.5 NA = 0.22 mounted in SMA-coupling, dismountable
  • Entrance slit
    70 µm x 1400 µm (optical entrance)
  • Grating
    Flat-field correction
  • Spectral range
    200 nm - 980 nm
  • Wavelength accuracy absolute
    < 0.5 nm
  • Reproducibility
    < 0.1 nm
  • Temperature - induced drift
    <0.01 nm/K
  • Spectral distance of pixel
    Δλ pixel ~ 0.8 nm
  • (FWHM: Full width half maximum)
    Δλ < 3 ...4 nm (UV-NIR Version)
  • Straylight
    0.1 % measured at 340 nm with deuterium lamp (transmission of NaNO2 solution, 50 g/l, 1 cm)
  • CCD Array
    Thermoelectrically cooled Hamamatsu S 7031
  • Number of pixels
    1044 x 64
  • Dimensions of pixels
    24 x 24 mm2
  • System data
    16 Bit AD conversion
  • Noise
    2...4 count standard deviation